Fluke MET/CAL Procedure ============================================================================= INSTRUMENT: BMI 155: (1 year) CAL VER /5520 DATE: 02-Sep-97 AUTHOR: Fluke Corporation REVISION: $Revision: 1.1 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 8 NUMBER OF LINES: 93 CONFIGURATION: Fluke 5520A STANDARD: 5500A/COIL ============================================================================= # # Compatibility: # 5500/CAL 5.0 or MET/CAL 5.0 or later. # # Subprocedures: # None # # Required Files: # None # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5520A accuracy file contains 90 day specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- R P F V 1.002 ASK+ K X 1.003 STD 5500A/COIL 1.004 HEAD PRELIMINARY INSTRUCTIONS 1.005 DISP Warm-up time: NONE 1.005 DISP Temperature: 0 - 50 degC 1.005 DISP Humidity: 0 - 95% RH 1.006 DISP [32] WARNING 1.006 DISP HIGH VOLTAGE is used or exposed during the performance 1.006 DISP of this calibration. DEATH ON CONTACT may result if 1.006 DISP personnel fail to observe safety precautions. 1.007 DISP Connect the 5520A to the UUT as follows: 1.007 DISP [32] 5520A NORMAL HI to UUT LEFT V INPUT 1.007 DISP [32] 5520A NORMAL LO to UUT RIGHT V INPUT 1.008 DISP Press the ON button. 1.008 DISP Press the VOLTS button (for 2 seconds). 1.008 DISP Press the RMS button (for 2 seconds). 1.009 HEAD {VOLTAGE} 1.010 DISP Press and HOLD the MODE button for 5 seconds, 1.010 DISP to set the unit to the freerun mode. 1.011 5520 1000 600V 1% 1/ 60H SI 2W 2.001 HEAD {HARMONIC VOLTAGE} 2.002 5520 60Vpp 60H SQ S 2W 2.003 DISP Press the HARM% button (for 2 seconds), and check the 2.003 DISP reading for harmonic #3. NOTE you will need to step 2.003 DISP through harmonic #1 and #2 before reading harmonic #3. 2.004 ASK- U 2.005 ACC 33.3pct TOL 2.006 MEMI Enter the UUT Harmonic #3 reading in percent: 2.007 MEME 2.008 MEMC pct 1.3U 3.001 ASK+ U 3.002 HEAD {DISPLAY FUNCTIONS} 3.003 HEAD DISPLAY FUNCTIONS: {WAVE} 3.004 5520 300V 60H SI S 2W 3.005 DISP Press the WAVE button (for 2 seconds). 3.006 EVAL Is the display a 60Hz sinewave? 4.001 HEAD DISPLAY FUNCTIONS: {GRAPH} 4.002 5520 60Vpp 60H SQ S 2W 4.003 DISP Press the GRAPH button (for 2 seconds). 4.004 EVAL Is the display a harmonic spectrum (GRAPH)? 5.001 HEAD DISPLAY FUNCTIONS: {THD} 5.002 5520 100V 60H SI S 2W 5.003 DISP Press the THD% button (for 2 seconds). 5.004 EVAL Is a reading of <1% DIST-IEEE displayed? 6.001 DISP Disconnect the 5520A from the UUT. 6.002 HEAD {CURRENT} 6.003 DISP Press the AMPS button (for 2 seconds). 6.003 DISP Press the WAVE button for 2 seconds. 6.003 DISP Press and HOLD the AMPS button until the 10A range 6.003 DISP is selected (as indicated in left edge of the bargraph). 6.003 DISP Press the RMS button (for 2 seconds). 6.004 DISP Loop a piece of heavy gauge copper wire from the 6.004 DISP 5520A 20A terminal to the AUX LO terminal. 6.005 DISP Connect the CURRENT PROBE to the UUT probe input. 6.005 DISP Clamp the current probe around the wire. 6.006 5520 10 5A 2% 2/ 60H SI HC 2W 7.001 DISP Unclamp the probe from the wire and remove the wire from 7.001 DISP the 5520A 20A and LO terminals. 7.002 DISP Connect the 5500A/COIL to the 5520A AUX HI and LO. 7.002 DISP Clamp the current probe around the Coil. 7.003 5520 500 100A 1% 1/ 60H SI T5 8.001 END